3. Specifications

3.1 General Specifications

Measurement function:

DC Voltage, AC Voltage, DC Current, AC Current, Resistance,

Frequency, Temperature, Capacitor, Continuity Check, Diode Test

Other functions:

Data Hold (D•H), Auto Hold (A•H), Range Hold (R•H), Maximum
value* (MAX), Minimum value* (MIN), Average value* (AVG),
Zero Adjustment (Capacitor, Resistance), Relative values, Save to

Memory*, LCD backlight.

*: For model KEW1052 only

Measuring method:

Σ modulation

Display:

4-digit (LCD)/7-segment

Maximum Reading: 6000

Polarity Indicator: “–” Appears automatically when the polarity is

negative

Overrange Indicator: “ OL ”

Low-battery Indicator: “

” Appears when the batteries become

low

Measurement cycle: 5 times per second

(except frequency measurement : one time per second, Resistance
measurement (6MΩ/60MΩ) : 2.5 times per second, capacitor
measurement (1000μF) : max.0.14 time per second)

Bar graph display approx 25 times per second (at AC, Ω)

Operating temperature and humidity ranges:

-10 to 55ºC, 80%RH or less (no condensation)

70%RH or less at 40 to 55ºC.

Storage temperature and humidity ranges:

-30 to 70ºC, 70%RH or less (no condensation)

Temperature coefficient:

(Accuracy at 23±5ºC× 0.1)/ºC should be added.

(Temperature ranges: -10 to 18ºC and 28 to 55ºC)

Power supply:

AA-size (R6/LR6) 1.5V batteries: 4

Battery life:

Approximately 300 hours

(Operating hours of alkaline batteries when in DC voltage-mode.)

Note: The battery life varies depending on the operating conditions.

Insulation resistance : 1000V DC, 100MΩ or more