SHIMPO CAG-2000 1 Channel User Manual

Page 21

Advertising
background image

19

‘

Press to determine the selected polarity. The next step is setting the

measuring range, and “RANGE” is shown on the alphanumeric display.

Fig. 9-3. Determining polarity.

(2) Setting measuring range

Changing the measuring range may call for the A/E converter to be switched over.

When changing the measuring range between 50 and 100

µm, the A/E converter

need not be switched over, but the instrument must be adjusted as described in “10.

Adjusting Zero Position and Sensitivity with Zero and Sensitivity Adjustment Knobs”.

In other cases, contact NIDEC TOSOK.



With “RANGE” shown on the alphanumeric display, press . The current

measuring range is shown. Press or to change the measuring range

to 20, 50, 100, or 200

µm.

Fig. 9-4. Changing measuring range.