10.2 Indication below mastering range

(1) When the indication does not reach the mastering range, the instrument is insufficient

in sensitivity. With the maximum master set in the measuring head, turn the

sensitivity adjustment knob clockwise so that the indication exceeds the maximum

master value by about 5 times the insufficiency from the maximum master value.

(2) Turn the zero adjustment knob clockwise to set the zero position at the maximum

master value.

Fig. 10-7. Adjusting for indication below mastering range.

(3) Set the minimum master in the measuring head. When the indication is about the

minimum master value, the adjustment is finished.

When the indication is not about the minimum master value, turn the zero adjustment

knob to bring the indication to the minimum master value. Set the maximum master

in the measuring head. If the indication does not come to about the maximum

master value, repeat steps (1) and (2) in 10.2 or 10.3.

ZERO

MAG

GAGE

ENT

RST

SET

MAS

C A G

2000

Turn to bring
indication to 100

Insufficiency

× 5 + Maximum master value

= 30

× 5 + 20

= 170
However, alphanumeric display cannot
show more than 100

1

3

S E T

M AS

2

Insufficiency
of 30

µm

Maximum
master value

ZERO

MAG

GAGE

ENT

RST

SET

MAS

C A G

2000

Turn to bring
indication to
maximum master
value

–10