29

’

Set the minimum master in the measuring head, and turn the zero adjustment

knob to set the indication at the minimum master value.

Fig. 10-3. Adjusting zero position.

“

Set the maximum master in the measuring head. If the indication is about the

maximum master value, the adjustment is finished.

Fig. 10-4. Finishing adjustment with zero and sensitivity adjustment knobs.

When the indication is not in the automatic mastering range (or the bargraph

display is illuminated red), refer to “10.2 Indication below mastering range” on

page 31 or “10.3 Indication above mastering range” on page 32.

ZERO

MAG

GAGE

ENT

RST

SET

MAS

CA G

2000

Set minimum master
in measuring head
and turn knob

1

3

S E T

M A S

2

Set indication at
minimum master
value

Adjustment
finished