2 external input – SHIMPO CAG-2000 1 Channel User Manual

Page 41

Advertising
background image

39

11.2 External input

(1) Description

Using external buttons and a foot switch, the instrument can:



Measure a work.



Reset itself.

‘

Be calibrated with the minimum master.

’

Be calibrated with the maximum master.

Note: Do not use the instrument through relay contact input.

(2) Preparation

The SW. ETC connector at the rear of the instrument is provided for external buttons.

It accepts a D-sub 15-pin (male) plug.

(3) Connector pin arrangement

Fig. 11-1. Pin arrangement diagram of SW. ETC connector at rear of instrument.

(4) Operation with external buttons



Measure button

(a) In the measure mode, press the measure button.

(b) The measured value is held. (This is called the hold measured value

condition.)

Note: The measured value cannot be held when the mastering result is NG.



RESET button

Clears the measured value hold.

Description

No.

1
2
3
4
5

GND

6

Measure button

7

RESET button

8

No.

Description

9

Maximum master button

10

Minimum master button

11

12
13
14
15