10.3 Indication above mastering range

(1) When the indication is above the mastering range, the instrument is excessive in

sensitivity. With the maximum master set in the measuring head, turn the sensitivity

adjustment knob counterclockwise so that the indication is returned from the

maximum master value by about 5 times the excess from the maximum master

value.

(2) Turn the zero adjustment knob counterclockwise to set the zero position at the

maximum master value.

Fig. 10-8. Adjusting for indication above mastering range.

ZERO

MAG

GAGE

ENT

RST

SE T

MAS

CA G

20 00

ZERO

MAG

GAGE

ENT

RST

SE T

MAS

CA G

20 00

1

3

SE T

M AS

2

42

Excess of
22

µm

Maximum
master value

Turn to bring
indication to –90

Maximum master value – (Excess

Ч 5)

= 20 – (22

Ч 5)

= –90

Turn to bring indication
to maximum master
value