15. Troubleshooting

Phenomenon

Cause

Remedy



Air supply pressure is not
stable.



Set regulator source
pressure at 300 kPa or
more.



Regulator is
malfunctioning.



Overhaul or change it.

‘

Measuring nozzle is worn. ‘ Change it.

Repeatability is not stable.

’

Piping and joints are
leaking.

“

Entry of moisture or oil
causes trouble to
instrument.

’

Check for leakage, and
retighten.

“

Clean instrument.



Air supply pressure is too
low or high.



Set regulator at 196 kPa.



Piping and joints are
leaking.



Check for leakage, and
retighten.

‘

Gap of measuring nozzle
is too small.

‘

Adjust to proper gap.

Zero adjustment knob is
ineffective.

’

Gap of measuring nozzle
is too large.

’

Adjust to proper gap.



Specified power is not
supplied.



Supply power of 85 to 264
VAC.



Zero position of air circuit
is improperly adjusted.



Calibrate instrument with
masters.

‘

Instrument is in hold
measured value mode.
Bargraph display is
illuminated green or red
(dark).

‘

Press to clear this
condition.

Bargraph display does not
operate.

’

Instrument is in set mode. ’ End set mode.



Power is not supplied.



Supply power of 85 to 264
VAC.



Fuse is blown.



Change it.

Bargraph and alphanumeric
displays are not illuminated.

‘

Power supply or internal
circuit is faulty.

‘

Ask NIDEC TOSOK for
repair.