11. I/O Description

11.1 Serial output (RS232C)

(1) Description

This instrument can output the measured values to a printer and communicate with a

personal computer through RS232C.

Command

Symbol

Hold measured value

E

Clear measured value hold

Clear master request

R

Request measured value

D

Minimum-mastering

N

Maximum-mastering

X

Change to Program 1

P01

Change to Program 4

P04

(2) Preparation

The RS232C connector at the rear of the instrument is provided for the personal

computer or printer. Connect the D-sub 9-pin (male) plug of an optional

communication cable to the RS232C connector. Three types of optional

communication cables are available for specific applications. For details, refer to

“13. Options”.

(3) Setting serial port

Baud rate:

9600

Bits/character: 8

Stop bit:

1

Start bit:

1

Parity bit:

None