SHIMPO CAG-2000 1 Channel User Manual

Page 42

Advertising
background image

40

‘

Minimum master button

(a) In the measure mode, set the minimum master in the measuring nozzle.

(b) When the bargraph display is stabilized, press the minimum master button.

(c) The instrument is calibrated with the minimum master and readied for

calibrating itself with the maximum master (this is called the maximum-

mastering ready condition).

’

Maximum master button

(a) In the maximum-mastering ready condition, set the maximum master in the

measuring nozzle.

(b) When the bargraph display is stabilized, press the maximum master button.

(c) The instrument is calibrated with the maximum master and returns to the

measure mode.

(d) When a mastering error occurs (refer to step “ in “7.1 Automatic mastering”),

use the control keys on the instrument to clear the mastering error.