DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

10

3.4 For several measurement tools connection cases

Shown here is the procedure up to measurement for cases wherein multiple measurement tools are

connected to 1 main body.

Please refer to [Chapter 3, 3.3

For first time measurement tool connection cases] when inputting of

settings for measurement tools that will be connected are not finished.

Replace the measurement tool

Turn ON compressed air

Turn ON power

Switch to program switching

mode

Program switching

Switch to master calibration

mode (detector adjustment)

Adjustment of detector

Switch to master calibration

mode (master calibration)

Master calibration

Measurement

Stopping of compressed air to the main body and replace
measurement tool.

Please refer to [Chapter 6, 6.2 Preparation] for details

Switch from measurement mode to program switching
mode
Please refer to [Chapter 6, 6.6 Program Switching] for
details

At program switching mode, reads the settings value and
master data of the measurement tool connected.
Please refer to [Chapter 6, 6.6 Program Switching] for
details

Switch to measurement mode after finishing program
switching.
Please refer to [Chapter 6, 6.6 Program Switching] for
details