SHIMPO DAG-2000 User Manual

Page 32

Advertising
background image

DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

32

10.3 Specifications

ITEM

SPECIFICATIONS

REMARKS

Input module

Air 1 Channel

Built-in AE2000 multi

No. of measurement items

1

No. of programs

10 PROG

1~10

100

50
20

Measurement range
[unit:

µm]

10

Set to each program
10

µm range is optional

5
2
1

0.5
0.2
0.1

0.05
0.02

Display resolution
[unit:

µm]

0.01

Set to each program





Can be selected except when range is 100
Can be selected when range is 10, 20
Can only be selected when range is 10

Main display
(Display color)

4 digits

(Red, green, orange, light dark)

Measurement value, judgment result

Multi-function display

40DOT/F.S.

Measurement value is displayed in analog
dots

Automatic master calibration

8 digits (Red)

Judgment result, settings values (English,
numbers)

Small range, large range

ZERO correction

Within

±50% of measurement range

Automatic master range

Sensitivity correction

Within

±50% of measurement range

And within 0.5 ~ 2.0X

Electric voltage

AC85~264V

Attached cable for AC100V

Frequency

50/60Hz common

Electric capacity

30VA

Dimension (L X W X H)
[unit:

µm]

120 x 180 x 150

During fitting of regulator
300 (depth) mm

Weight

2.2kg

2.8kg (with regulator)

Operating temperature

0~45

°C

Push button input
Foot switch

4 points

Measurement command, RESET, master
calibration (2)

Serial communications

RS232C 1 port

Output of measurement value and
judgment result

High precision measurement

Measurement range 10

µm

Digimatic output

1 port

Attached DP-1 connection cable

Input signal
8 points

Measurement command, Reset

Master calibration

Program switching

READY

Master OK

Common signal at 1, 2

Judgment
output

BCD output

Output signal
Open corrector
24 points

1. Rank output

2. BCD output

1, 2 or either one

Single type

Measurement command

Foot switch

2 consecutive type

Measurement command, RESET

Serial communications cable

For RS232C

Peak measurement function

+PEAK, -PEAK,

TIR(=+PEAK-(-PEAK)),

TIR/2,

(+PEAK+(-PEAK)/2

Based on measurement value change
With auto measurement function

Rank extension function

99 ranks maximum

OK range

Regulator

Precision

regulator

Filter

Op

tio

n

al

Compressed air purifier

When there is too much moisture and oil