DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

15

CHAPTER 5

MEASUREMENT

Measurement is possible when [MEAS] mode lamp is lit and LED of [MAS] is unlit.

If [MAS] LED is blinking, this means that master calibration has not been performed so please

perform first [Chapter 4 Master Calibration].

5.1 Measurement value display

Display contents for the main display and English display can be changed by setting it.

Display device

Display contents

Settings item

name

Settings Remarks

Measurement value
[

µm]

MainDisp

MEASURED

µm

Displays the measurement
value in

µm

Measurement value
[mm]

Ditto

MEASURED mm

Displays the measurement
value in mm

Judgment result

Ditto

JUDGMENT

-OK=1, OK=2, +OK=3,
-NG & +NG=no display

Main display

Not used

Ditto

NO USE

Measurement value
[mm]

CharDisp

MEASURED

Displays the measurement
value in mm

Judgment

result

Ditto

JUDGMENT

The left 3 digits is the
program number.
The right 4 digits is the
judgment result

English display

Plain bar

Ditto

BAR

Position display of
measurement value within
the measurement range

Please select in settings item [RESOLUTION] for the resolution function in the measurement values display.

5.2 Display color of main display

Display color of the main display changes depending on the judgment result.

If judgment result is:

OK, the color will be GREEN.

-OK and +OK, color will be ORANGE,

-NG and +NG, color will be RED.

5.3 Measurement value hold

If ENT command is inputted during measurement, measured value will be on hold (saved) and if

external I/F is being used, measurement value and judgment results will be outputted.

Also, display color will change from light to dark.

To cancel measurement value saving, RST or input RESET.

Note: Cannot hold (save) when master calibration in NG.