DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

9

3.3 For first time measurement tool connection cases

Shown here is the procedure up to measurement wherein measurement tool is connected to the

main body for the first time.

Note: There are cases wherein detector adjustment is not needed depending on the measurement

tool.

Turn ON compressed air

Turn ON power

Switch to settings mode

Change the settings

Switch to master
calibration mode

(detector adjustment)

Detector adjustment

Switch to master

calibration mode (master

calibration)

Master calibration

Measurement

Switch from measurement mode to settings mode
Because of work tolerance change, change the judgment
limit values.
Settings value change can be done in setting mode.

Please refer [Chapter 6, 6.4 Settings] for details

At settings mode, input the settings value for the
measurement tool that will be connected.
Please refer to [Chapter 6, 6.4 Settings] for details

Switch to measurement mode after finishing settings
change
Please refer to [Chapter 6, 6.4 Settings] for details

Switch from measurement mode to detector adjustment
Please refer to [Chapter 6, 6.5 Detector Adjustment] for
details

Switches from detector adjustment to master calibration
Refer to [Chapter 6, 6.5 Detector Adjustment] for
details.

Adjustment of detector through ZERO and sensitivity
tuner.
Please refer to [Chapter 6, 6.5 Detector Adjustment] for
details