DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

14

4.3 Clearing of master data

Shown here is the clearing method of ZERO correction and sensitivity correction data
Ordinarily not used. Please use for wear check of measurement tool.

4.4 Master calibration error

In case of master calibration error, the following error details will be shown in the English display:
“ERR ZERO”

ZERO correction is out of range

“ERR MAG”

Sensitivity correction is out of range

“ERR REV”

Data during ZERO correction is smaller than data during sensitivity
correction

In case the above errors are displayed, it is necessary to adjust through the ZERO/sensitivity
adjustment tuner.

Please refer to [Chapter 6, 6.5 Detector Adjustment] for details.

Measurement mode

Switch to master
calibration mode

(master calibration)

Switch to master
calibration mode

(detector adjustment)

Switch to master
calibration mode

(Clears master data)

Clear master data

Switch to master
calibration mode

(master calibration)

Please confirm that it is in measurement mode (mode lamp [MEAS]

will light up).
Please refer to Chapter 5 Measurement for details

Press /

MAS

continuously for 2 seconds.

Will switch from measurement mode to master calibration mode

(master calibration)

After “MIN M.” or “ZERO M.” has been shown in the English

display, mode lamp [MAS] with orange color with light up.
Please press /

MAS

.

Switches from master calibration mode (master calibration) to master
calibration mode (detector adjustment)

“ADJ.” will be shown in the English display.

Please press /

MAS

.

Switches from master calibration mode (detector adjustment) to master
calibration mode (master data clear)

“MST CLS.” will be shown in the English display.

Please press ENT.
Master data will be cleared.
After “M. CLS OK” has been shown in the English display, mode
lamp [MAS] with orange and red colors with light up alternately.

Returns to condition .