DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

29

7.2 External button input

(1) Outline

This device allows the connection of the non-electrical contact of the external button or foot switch
to the [SW, ETC] at the back of the main body.
Measurements, RESET, large range master calibrations, and small range master calibrations can be
performed.


(2) Preparation

[SW. ETC] at the back of this device is the connection port to the external button.
D-sub 15P (OSS) can be connected.

(3) Switch input connector

External button and also non electric contacting foot switch can be connected.
Input can be activated by short circuiting to the ground pin (6).
Please use D-sub 15P (OSS).

Note:

Please set the cable length within 2m.

Cannot use other than push button or foot switch.

Please use DS I/O base (optional) for the connection of sequencer and relay.
Please do not connect to the pin number that has no signal name. This becomes cause of
accidents.












(4) Movement based on external button
a. Measurement button

(1) At measurement mode, please press [Measurement Button].

Outputs the measurement value to external machine

Note: Cannot output when master calibration is NG

(2) At settings item [MEAS SW],

If [HOLD] is selected, measurement value will be on hold (saved), and data will be outputted.
If [NOT HOLD] is selected, data will only be outputted.

b. RESET button (Reset)

(1) Cancels hold.

c. Large range master calibration

(1) At measurement mode, please set the large range master to the measurement tool.
(2) Please press the [Large range master calibration button] when measurement value stabilizes.

Returns to measurement mode after performing large range master calibration.

Note: The above movement will not be performed when [ZERO M.] at settings item

[CalMode] is selected.

d. Small range master calibration

(1) At measurement mode, please set the small range/ZERO master to the measurement tool.
(2) Please press the [Small range master calibration button] when measurement value stabilizes.

Performs small range/ZERO master calibration and then returns to measurement mode.

1

GND 6

Measurement 7

RESET 8

9 Large range master calibration
10 Small range and ZERO master calibration




15