SHIMPO DAG-2000 User Manual

Page 19

Advertising
background image

DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

19

(3) Settings details

This explains the settings item name and settings details at the settings mode.
Settings mode is mainly divided into the following 3:

• Normal

[Standard settings] --- master values, judgment limit values

can be set

• Detail

[Detailed settings] --- measurement range, display analysis function, polarity

ditto

• System

[System settings] --- display data, moving average

ditto


① Settings item names and settings details at Normal [Standard settings]

Settings concerning master values

• [MIN M.] --- Please input the small range master values used in master calibration (2 masters).

• [MAX M.] --- Please input the large range master values used in master calibration (2 masters).

• [ZERO M.] --- Please input the ZERO master values used in master calibration (1 master).

Settings concerning judgment limit values

• [-NG/-OK] --- Please input the limit values of –NG and -OK

• [-OK/ OK] --- Please input the limit values of –OK and OK

• [ OK/+OK] --- Please input the limit values of OK and +OK

• [+OK/+NG] --- Please input the limit values of +OK and +NG
In case judgment of –OK is not needed, please set the same settings values to [-NG/-OK] and

[-OK/OK] and in case judgment of +OK is not needed, please set the same settings values to
[OK/+OK] and [+OK/+NG].

Settings concerning master

• [CORRECT] --- Machine difference correction values

can be set

Adds and displays the above data to the master correction data.

② Settings item name and settings details at Detail [Detailed settings]

• [RANGE] --- Measurement range

can be selected

10

µm is optional.

• [RESOLUTION] --- display analysis function

ditto

Settings concerning detector

• [POL] --- Polarity

Please select + for inner diameter measurement and – for outer dimension measurement

• [GAIN] --- Sensitivity rough adjustment values

can be set

Normally, settings values is fixed depending on the selected measurement range.
Please input:
24 if measurement range 100

µm is selected

33 if measurement range 50

µm is selected

68 if measurement range 20

µm is selected

(204 if measurement range 10

µm is selected)

In case measurement range is changed, the above values will be set.

• [CONSTANT] --- Sensitivity fine adjustment constant can be set

Normally, please input 1.000.

Settings concerning master

• [CalMode] --- Master calibration method

can be selected

Please select:
[MIN & MAX] for master calibration by small range or large range master.
[ZERO M.] for master calibration by ZERO master.

③ Settings item name and settings details at System [System settings]

• [MEAS SW] --- External button input movement

can be selected

• [SMOOTH] --- Moving average

ditto

Settings concerning display

Data shown in:

• [MainDisp] --- Main display

can be selected

• [CharDisp] --- English display

ditto

Advertising