3 viewing the results – Techne PrimeQ User Manual

Page 158

Advertising
background image

158

End-point: Uses the last reading only (we recommend that >1 reading is used for

accuracy).

All readings (default): Averages all readings in the stage

Last readings: Averages a user-specified number of last readings (the default is 5, or 1 if

there are less than 5 readings)

Specify range: Specify a range of readings to be averaged that best suit the data (the

default is the last reading).

3.18.2.2 Report Options and Summary

• Click Next to lead through to the Report Options screen. This allows the user to choose

whether to include the multi-read table in the report. Check the box to select.

• Click Next to view a Summary of the setup.

• Click Back to change any settings or Cancel to abort the procedure.

• Click Finish to complete the set up.

3.18.3

Viewing the results

During the run, the real-time collection of data can be monitored on the Run Screen. The plate
layout shows the fluorescence curve on a per-well basis and the temperature profile plot indicates
how far the run has progressed.