Techne PrimeQ User Manual

Page 63

Advertising
background image

63

Run information: The user can enter a user name and add any other comments relevant to

the program.

Instrument settings: User-defined settings for heated lid temperature, disabling of the

instrument buttons, plate type, sample volume and whether to read entire plate or include a
final hold in the PCR program.

Program file view: Displays the structure of the program thereby providing the user with a

clear view of the overall protocol.

Temperature profile: Provides a summary plot showing the temperature profile for each

cycle of the stage (or the complete program when viewing at the Program level).

Clicking on the padlock icon to a

locked position prevents the user from

accidentally saving over the existing file – to save the changes the user must
use the

File menu Save As… option and save the file under a different name.

3.4.3.2

Run information

• Enter a user name and any comments in the appropriate text boxes (optional).

3.4.3.3

Instrument settings

• Define the general instrument settings for the run.

Advertising