Blue-White CHEM-PRO C3 ProSeries Diaphragm Pump User Manual
Page 14
Page 14
Page 15
Chem-Pro
Chem-Pro
®
®
Flow verification sensor sold separately.
Flow verification system is designed to stop pump in an event sensor does not detect flow during pump operation.
Indicating an empty chemical tank, clogged injection fitting, loose tubing connection, etc.
To allow pump to clear any gasses that may have accumulated over time, an alarm delay time value from 1 to 255
seconds must be programmed.
Note: An alarm delay of 000 seconds disables FVS system.
Step 1
Ensure pump is stopped and LCD reads “OFF.”
Note: Mode cannot be changed while pump is in running.
Press and release STOP button if pump is running.
Press and release MODE button multiple times until Mode 0
is selected.
Step 2
With Mode 0 selected, press and hold MODE button until
‘Remote’ icon begins flashing.
This indicates that you’ve entered Setup menu.
Step 3
Press and release MODE button to scroll through menu until
you see FVS icon.
If you pass FVS screen, continue to press and release
MODE button until FVS icon appears.
Step 4
FVS icon will appear for 1 second, followed by numbers.
These numbers indicate delay time setting for FVS.
Select a delay time in seconds. Delay time is amount of time
pump will wait to receive a pulse from sensor until an alarm
it triggered.
A delay time of 00 deactivates FVS feature.
Mode 0
SEt
0
Mode 0
FVS
0
Mode 0
no
0
Remote
Icon
Mode 0
00
0
Step 5
After you’ve made your selection, press and release MODE
button. This saves your setting.
You can now modify other settings in Mode 0 or you can exit
Setup by; press and hold MODE button for a few seconds
until you return to Run screen.
Mode 0
set
0
REMOTE
SEC
7.2
Mode 0 - Set FVS (flow verification system)
7.2
Mode 0 - Set FVS (flow verification system) - Continued
Flow Verification Sensor is designed to give you two installation options.
Sensor can be installed:
!
Directly onto pump head of Chem-Pro® pump, discharge side.
!
Anywhere on discharge side of Chem-Pro® pump.
Wiring for sensor can be connected directly to a Chem-Pro® pump. Pump will stop pumping if sensor detects no
flow. A relay will then close allowing for remote alarm indication or initiation of a back-up pump.
Install FVS Flow
Sensor - Flow Verification Sensor should be installed on inlet (suction) side of pump tube.
When installing directly onto pump 3/8” tube discharge fitting:
Sensor includes a PVC tubing insert, located inside sensors female thread connection, that is designed to
seal sensor onto pump tube adapter. Thread sensor onto pump tube until tubing insert is snug against pump
tube fitting - do not over-tighten.
Sensor Model Number
Actual Working Range with
Chem-Pro® Pump
ML/Min
ML/Min
30-200
FV-100
30-300
Published Flow Range
50-900
FV-200
100-1000
100-1800
FV-300
200-2000
300-3000
FV-400
300-3000
500-5000
FV-500
500-5000
700-7000
FV-600
700-7000
Confirm FVS flow range - Flow Verification Sensor (FVS) will only function within its operating range. See chart
for available ranges.
NOTE: If pump output is less than 30 ml/min, sensor will not detect chemical and a signal will not be sent to
pump, resulting in an alarm condition.
NOTE: For low viscosity (water-like) fluids only. Consult factory if attempting to use with viscous fluids.
C3V | Diaphragm Metering Pump
Made in the USA
MODE
PRIME
Program
Minimum
Maximum
Blue-White Ind.
IP66
NEMA 4X
FVS
Sensor
Time-out - Chem-Pro® pumps have a time-out setting of approximately 20 seconds while in configuration menu. If built-in timer
exceeds 20 seconds without a button being pressed, then pump will exit configuration menu. Changes will automatically be saved if
programming mode is allowed to Time-Out.