Mpcal= (multipoint calibration) – PNI TCM 2.6 User Manual

Page 30

Advertising
background image

PNI Sensor Corporation

DOC#1009269 r11

TCM2.5 & TCM2.6 User Manual

Page 30 of 49

mpcal=

(Multipoint Calibration)

Description

Initiates a multipoint calibration. This command is only valid when used in

conjunction with Enter Continuous Sampling Mode (go). The TCM2.5/2.6 samples its
sensors and adds the data point to the current set of calibration data.


Syntax

mpcal=e<cr>

:<cr><lf>

go<cr>

.......

...... take data in continuous mode

h<cr> (or h)

:<cr><lf>

mpcal=d<cr>


Valid Values

Input

Description

e

e, enable multipoint calibration

d

d, disable multipoint calibration (default setting)

Table 8: Valid

mpcal= Values

Output

HnVnMn.nn<cr><lf>

The TCM2.5/2.6 only reports a calibration score after it has been put in Standby mode (h), and

multipoint calibration has been disabled (mpcal=d)

Example

mpcal=e<cr>

:<cr><lf>

go<cr>

........

....... take data in continuos mode

h<cr>

:<cr><lf>

mpcal=d<cr>

H4V3M5.00:<cr><lf> (This is an example of a poor score)

Query Response

mpcal?

:mpcal=

Related Commands

lc? (Query Last Calibration Score)