Timeconst= (set time constant for digital damping) – PNI TCM 2.6 User Manual

Page 41

Advertising
background image

PNI Sensor Corporation

DOC#1009269 r11

TCM2.5 & TCM2.6 User Manual

Page 41 of 49

timeconst=

(Set Time Constant for Digital Damping)

Description This command is used with the digital damping command. It sets the time constant to

nn samples for the digital damping.

Syntax

timeconst=n<cr>

Output

:[Ennn]<cr><lf>

Valid Values

Input

Description

Integer value between 1

– 4

Set the measured time constant = 4

Integer value between 5

– 8

Set the measured time constant = 8 (default setting)

Integer value between 9

– 16

Set the measured time constant = 16

Integer value between 16

– 32

Set the measured time constant = 32

Table 21: Valid

timeconst= Values

Example

timeconst=16<cr>

!Filter over 16 measurements

:<cr><lf>

Query Response

timeconst?

:timeconst=

Related Commands

damping= (Set Digital Damping)

Advertising