7. CALIBRATION

This function adjusts the scale (indicator) for accurate weighing.
Calibration must be done when the FS-D is initially connected to a load cell (base unit).
Calibration may also be required according the changes in the environment.

‰

When the scale (indicator) has been moved.

‰

When the ambient environment has

significantly changed.

‰

For regular calibration.

Loose the lock screws on the rear side of the
display pod, and remove the (CAL) switch
cover. The calibration switch inside.


7-1. Calibration using a weight

1. Warm up the indicator with a weighing platform connected for at least half an hour.

‰

Change the user function setting “f01” or place something on the platform to

disable the auto power-off function.

‰

When turning the power on and the indicator may display ------- ,

press the

PRINT/ENT

key to proceed to the weighing mode.

2. Press and hold the calibration (CAL) switch until

Cal

appears, and release the switch.

‰

Press and hold the

0

key for about 5 seconds, then you

can also enter the calibration mode

(when “C01-0” is set).

‰

Press the

ON/OFF

key or (CAL) switch to exit without

calibration.

3. Press the

PRINT/ENT

key, then the gravity

acceleration value will be displayed.

‰

It is not necessary to set the gravity acceleration value

when calibrating the indicator with a calibration weight
at the place where it is to be used. (See the next
section about the gravity acceleration correction.)

4. Press the

PRINT/ENT

key, then

Cal

0

will be

displayed.

5. Make sure that there is nothing on the platform, and wait

until the STABLE indicator turns on.

6. Press the

PRINT/ENT

key. The indicator inputs the

zero point, and the display shows

5pn 1

and the

weight value for calibration (SPAN calibration).

‰

The weight value is equal to the capacity. When you

enter with “kg” or “g” mode, then the value is “kg”.
Entering with “lb” or “oz”, then “lb”.

-0

5pn

0

1

0

-00

Cal

00

-00

9.7985

-00

Cal

0

0

-000

30.00

CAL switch cover

Calibration (CAL) switch