A&D FS-D User Manual

Page 8

Advertising
background image

6

Setscrew B

Setscrew A

4. Solder the wires to the pin contacts of the connector.
‰

A 5 m or less load cell cable is recommended as the

FS-D uses a 4 wire system (no remote sensing).









5. Reassemble the connector plug

and tighten the setscrew A.

6. Move the load cell cable back

and forth and around, and tighten
each parts again. Finally, tighten
the setscrew B.

7. Connect the connector plug to the

load cell connector on the bottom
of the display pod.

4-3. Verifying Load Cell Output and Input Sensitivity

The input sensitivity is 0.2 µV/division or more. When designing a weighing scale, the
load cell output voltage should satisfy the input sensitivity of the indicator. Or the scale
should be designed so that the following calculating formula is satisfied.

‰

In order to achieve a scale with stable performance, select the input sensitivity as

large as possible.

Calculating formula:

A : Rated capacity of load cell (kg)
B : Rated output (mV/V)
d : Min. display (smallest of kg) / scale division

E : Excitation voltage (mV) = 5000 mV for FS-D

n : Number of load cells

Example of calculation:

In the case of designing a scale with a capacity of 60 kg, using a load cell with a
rated capacity of 100 kg and rated output of 1 mV/V:

Load cell capacity:

A = 100 kg

Rated output:

B = 1 mV/V

Set function setting C02-3, then
Min. display (smallest):

d = 0.005 kg

Excitation voltage:

E = 5000 mV

Number of load cell:

n = 1

n

A

d

B

E

Ч

Ч

Ч

0.2

2

.

0

25

.

0

1

100

005

.

0

1

5000

>

=

Ч

Ч

Ч

SIG-

EXC+

SIG+

EXC-

Shield

Load cell

Load cell connector

Load cell cable