Circuit diagram – A&D FS-D User Manual

Page 19

Advertising
background image

17

Command H I , + 0 0 1 0 0 C

R

L

F

Assume 2 decimal place number.

Reply

H I , + 0 0 1 0 0 C

R

L

F

1.00% will be set as LO limit %.

Command L O , + 0 0 1 0 0 C

R

L

F

Assume 2 decimal place number.

Reply

L O , + 0 0 1 0 0 C

R

L

F

1.00% will be set as LO limit %.


‰

Store the comparator limits into the specified memory number. (No reply for f20-1.)

Reply

H I , 0 1 , + 0 0 1 2 0 0 , + 0 0 0 9 0 0 C

R

L

F

Reply

M L , 0 1 , + 0 0 1 2 0 0 , + 0 0 0 9 0 0 C

R

L

F


9-3. OP-04 RS-422 / 485

The followings are the corrections and the additional information for the optional
interface. (“3-2. OP-04 Specifications” in 1WMPD4001368).

Circuit diagram

The frame ground and the shield are connected internally and there is no difference
between them in use.

Multi-connection with Print key mode (new function)

This is one of the applications when a PC and more than one FS-D are connected.
The indicator will prepare the weighing data by pressing the

PRINT

key first and will

send out the prepared data after receiving a command from the PC.

1. Set the Function “f06-5” for all of the indicators.

2. Weigh something and wait for the STABLE indicator to turn on (the indicator having

f18-##).

3. Press the

PRINT

key, the indicator will temporarily store the weight data and turn

the PRINT indicator on.

‰

The indicator does not accept the

PRINT

key while the PRINT indicator on.

4. Send the “@##S” command to the indicators from the PC.
5. The indicator (f18-##) will respond to the command to send the weight data in

memory and turn the PRINT indicator off.

The address ## = 23 (f18-23).
‰

A request for weight data.

Command @ 2 3 S C

R

L

F

Reply

@ 2 3 S T , + 0 0 1 2 . 3 4 5 _ k g C

R

L

F

Reply

@ 2 3 I C

R

L

F

The indicator does not have any stored data to send.

‰

While the PRINT indicator is on, the indicator will send back “@##I” to commands

other than “@##S” command.

5 Frame ground FG
6 Shield

5 Shield
6 Frame ground FG

Advertising