A&D FS-D User Manual

Page 15

Advertising
background image

13

‰

If you need ZERO calibration but do not need SPAN

calibration, turn the power off to exit from the calibration
procedure.

7.

To calibrate with a different weight, change the
displayed value using the 10-key pad.

‰

If you enter the wrong number, press the C key. The

value returns to the capacity and enter the number again.

8. Place a calibration weight on the platform with the same

value as displayed, and wait until the STABLE indicator
turns on.

9. Press the PRINT/ENT key. The indicator calibrates

SPAN and

end

will appear. Remove the weight

from the platform, and turn the power off.

‰

Note

The value set in step 7 will be cleared after the power is switched off.
If the indicator will be moved to another location, set the gravity
acceleration value for the current location and calibrate the indicator
according to the procedure above. See the next section to set the value.



7-2. Gravity acceleration correction

When the indicator will be moved to different place from the calibration location, it
should be calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the indicator. To use this function, set the gravity acceleration value to the
value of the location where the calibration will be done. Refer to the gravity
acceleration map appended to the instruction manual of the FS-i series.

1. At step 3 of the previous section “7-1. Calibration using

a weight”, enter a new value using 10-key pad.

‰

The integral part “9” is fixed, enter the value after the

decimal point.

‰

If you hit the wrong number, press the

C

key. The

value returns to the original number. Enter the number
again.


2. Press the

PRINT/ENT

key, then

Cal 0

will be

displayed.


3. Go to step 5 of “7-1. Calibration using a weight”.

-000

20.00

-00

end

00

-00

Cal

0

0

-00

9.8050