Additional information, 1. f-function list – A&D FS-D User Manual

Page 17

Advertising
background image

15

9. ADDITIONAL INFORMATION

This section describes the additional functions and corrections to the instruction manual of
FS-i series (WM+PD4001332 and 1WMPD4001368) which are common to the FS-D.

9-1. F-Function list

The following function settings are added (“11-2. Function list” in WM+PD4001332).

Item

Function number

Description

f06- 5

Multi-connection with print key mode

(RS-422/485)

f06- 6

Auto-print mode + data & OK

Serial interface
Data output
mode

f06- 7

Auto-print mode +/- data & OK

The UFC format
is applicable to
f06-2 to 4.

f13- 0

Zero tracking OFF

Š

f13- 1

Zero tracking ON, 0.5d/sec

f13- 2

Zero tracking ON, 1.0d/sec

Zero tracking

f13- 3

Zero tracking ON, 2.0d/sec

Setting C04-0,
f13-2 and 3
cannot be set.

Š

f21- 0

Auto-tare function disabled.

Auto-tare
function

f21- 1

Auto-tare function enabled.

f22- 0

Immediately after OK and stable

f22- 1

0.5 sec. after OK and stable

Š

f22- 2

1.0 sec. after OK and stable

f22- 3

1.5 sec. after OK and stable

f22- 4

2.0 sec. after OK and stable

f22- 5

2.5 sec. after OK and stable

f22- 6

3.0 sec. after OK and stable

f22- 7

4.0 sec. after OK and stable

f22- 8

5.0 sec. after OK and stable

Auto-tare timing

f22- 9

6.0 sec. after OK and stable

Timing to tare
automatically
after the comp.
OK and weight
stable.
To be used with
f21-1.

Š

f23- 0

Function disabled

Auto-tare for the
initial weight

f23- 1

Tares the initial (container) weight.

Automatic

operation.

Š

f24- 0

Normal comparison

Normal/Negative
comparison

f24- 1

Negative comparison for take-away

Š

Factory setting


9-2. OP-03 RS-232C

The followings are the corrections and the additional information for the optional
interface (“2-2. OP-03 Specifications” and “2-3. Command Mode” in 1WMPD4001368).

Circuit diagram



4 Signal ground SG
5 Frame ground FG
6 Shield

4 Signal ground SG
5 Shield
6 Frame ground FG

Advertising