Specifications, 1. capacity and minimum display, 2. specifications – A&D FS-D User Manual

Page 21

Advertising
background image

19

11. SPECIFICATIONS

11-1. Capacity and minimum display

C02- setting

0* 1 2 3 4 5 6 7

Capacity 6

15

30

60

150

300

600

1500

0.002* 0.005

0.01 0.02 0.05 0.1

0.2

0.5

0.001 0.002

0.005

0.01 0.02 0.05 0.1 0.2

kg

Min. display

0.0005 0.001

0.002

0.005

0.01 0.02 0.05 0.1

Capacity 6000

15000

30000

2* 5 10

1 2 5

g

Min. display

0.5 1 2

NA NA NA NA NA

Capacity 15

35

70

150

350

700

1500

3500

0.005* 0.01 0.02 0.05 0.1 0.2 0.5

1

0.002 0.005

0.01 0.02 0.05 0.1 0.2 0.5

lb

Min. display

0.001 0.002

0.005

0.01 0.02 0.05 0.1 0.2

Capacity 240

560

1120

2400

5600

11200

0.1* 0.2 0.5 1 2 5

0.05 0.1 0.2 0.5 1 2

oz

Min. display

0.02 0.05 0.1 0.2 0.5 1

NA NA

Capacity 15

35

70

lb-

oz Min. display

0.1

0.1

0.1

NA NA NA NA NA

*: Factory setting

NA: Not Available

11-2. Specifications

Input sensitivity

0.2 µV / d (d: the smallest Min. display of kg)

Input signal range

0 mV ~ 13 mV

Load cell excitation

voltage

5 V DC ±5% (no remote sensing)

Load cell drive capacity

Up to 4 x 350 ohm load cells

Zero

±(0.2 µV ±0.001% of zero adjustment voltage) / °C (typ.)

Temperature

coefficient

Span

±0.001% / °C of reading (typ.)

Non-linearity

±0.01 % of full scale

Display

7 segment LCD display (Character height 18.6 mm) with backlight

60 segment analog sweep display with backlight

Display update

Approximately 20 times per second

Operating temp.

-10°C ~ 40°C / 14°F ~ 104°F, less than 85% R.H.

Power supply

AC main (100V ~ 240 V, 50/60 Hz) or SLA Battery (option)

Dimension

250 (W) x 202 (D) x 266 (H) mm / 9.8 (W) x 8.0 (D) x 10.5 (H) in.

Weight

Approximately 2.7 kg

Advertising