AirSep 20626704-C1 User Manual

Page 11

Advertising
background image

Eclipse User Manual — 11

DC Power Supply

Item# 5942-SEQ

DC POWER SUPPLY

A DC Power Supply allows the system to operate from DC

outlets, such as those found in motor vehicles.
1. Start your vehicle.
2. Insert the DC cord into the recessed power receptacle on the

side of the device.

3. Insert the DC Power Supply plug into the DC power outlet

in your motor vehicle.

When the device is properly connected and receiving power from

the DC power source,a green indicator light on the Power Supply

and the External Power Indicator on the control panel will

illuminate.

Eclipse 5 DC Supply Connection

When the Eclipse 5 is connected to the DC supply in a vehicle,

the following applies:
• The Eclipse 5 will run at 3.0LPM on continuous flow while

being operated with the DC Supply in a car.

• The Eclipse 5 will charge the battery while on DC Supply in a

car while running on continuos flow at 2LPM or less.

• If the Eclipse 5 battery is being charged, the battery symbol

will waterfall. If the Eclipse 5 battery is not charging, the

battery symbol will be displayed solid.

Settings Available When Operating on DC for Eclipse 5

Flow Setting

Continuous

Bolus Size (mL)

Pulse

0.5

Yes

-

-

1.0

Yes

16

Yes

1.5

Yes

24

Yes

2.0

Yes

32

Yes

2.5

Yes

40

Yes

3.0

Yes

48

Yes

-

-

56

Yes

-

-

64

Yes

-

-

72

Yes

-

-

80

Yes

-

-

88

Yes

-

-

96

Yes

-

-

128

Yes

-

-

160

Yes

-

-

192

Yes

Note: While operating the device from a DC power supply, the Battery may
not charge.
Note: When operating the Eclipse from a DC power supply, it is recom-
mended that the Pulse Dose Mode be used if prescribed to use a pulse
dose device.
Note: To ensure that the device is utilizing the DC power, check that the
External Power Indicator is illuminated. If a beep is heard, this could indi-
cate a loss of external power. Check if the green Battery Power Verification
LED is on or off. If the battery is removed on DC power, it will ensure no
loss of charge while the unit is running. If the indicator is not illuminated,
the battery is powering the device.

WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE
AIR OUTLET vENTS TO BECOME BLOCKED. THIS CAN CAUSE
THE OXYGEN CONCENTRATOR TO OvERHEAT AND IMPAIR

PERFORMANCE.
WARNING: IF THE vEHICLE’S DC POWER SOURCE DROPS BELOW 11.5
vOLTS, THE ECLIPSE WILL REvERT TO THE BATTERY OPERATION (IF
THE BATTERY IS PRESENT AND CHARGED). (THE EXTERNAL POWER
INDICATOR MAY REMAIN ON.)

Note: Actual performance of the Eclipse may vary depending upon usage
conditions of the motor vehicle.
Note: When operating with a DC Power Supply, Eclipse 3 Continuous Flow
settings of 2.5 and 3.0 LPM and Pulse Dose bolus sizes of 128, 160, and
192 mL are not available.
Note: When operating with a DC Power Supply, Eclipse 3 settings greater
than 2.0 LPM Continuous Flow, will automatically switch to the 2.0 LPM
Continuous Flow setting.

Advertising