First time eclipse operation – AirSep 20626704-C1 User Manual

Page 14

Advertising
background image

14 — 20626704 C (English)

Step 3: Connect the Oxygen Supply Tubing or Nasal
Cannula to the Oxygen Outlet

Replace the cannula and oxygen supply tubing regularly, as recommended by
your home care provider. Consult your home care provider for cannula and
supply tubing cleaning and replacement information.

Note: Ensure the cannula is fully inserted and secure. This ensures that the
Oxygen Concentrator can properly detect inspiration for oxygen delivery.
During inhalation, you should hear or feel oxygen flow to the prongs of
the nasal cannula.

Step 4: Select the Flow Mode

Press the Flow Mode button to select the desired mode. Pressing

this button repeatedly will toggle you back and forth between

Pulse Dose Mode and Continuous Flow Mode. The proper

mode and proper setting should be prescribed for you by your

physician.

Continuous Flow Mode Operation:

When operating in the Continuous Flow Mode, a continuous

supply of oxygen measured in liters per minute (LPM) will flow

through your tubing and nasal cannula.

Pulse Dose Mode Operation:

When operating in Pulse Mode, a bolus of oxygen, measured in

milliliters (mL), is delivered upon inspiration. DO NOT use a

humidifier in the Pulse Dose Mode.

To ensure proper saturations, secure a prescription for pulse dose

from your physician with consideration to your needs at rest,

during exercise, and when traveling at altitude. Contact your care

provider for prescription and boarding forms.

When operating in pulse dose mode, the Eclipse supplies a

continuous flow of oxygen if a breath is not detected. The device

will continually check for an inspiratory effort every 15 seconds.

Once an inspiration is detected, the device reverts to measured

bolus delivery. To exit Pulse Dose Mode and engage Continuous

Flow Mode, push the flow mode button.

First Time Eclipse Operation

Step 1: Positioning Your Eclipse for Use

Place the Eclipse in a well-ventilated, well

lit area. Be sure the air inlet and exhaust

vents are not obstructed.

Position the Eclipse so that all audible and

visual indicators or alarms can be easily

seen and heard.

Be sure the air inlet filter is in place before

operating your Eclipse. If the air inlet

filter is missing, contact your home care

provider. If the air inlet filter is dirty, wash

with warm soapy water, rinse with clear

water and allow it to dry before placing

the air inlet filter back in the device. If

necessary, replace it with a new, clean air inlet filter.

Plug the device into a grounded AC Power outlet, a DC Power

source, or be sure there is a fully charged Battery installed.

Step 2: Power ON the Device and Allow it to
Warm-Up

Press and hold the “ON/OFF” Button for two (2) seconds to

power ON your Eclipse. A brief audible and visual power-on self-

test will occur. Verify that all the power verification indicators

are illuminated and that a buzzer sounds to indicate the device is

operating properly.

Exhaust vent

Air Inlet Filter

The Eclipse has an oxygen concentration status indicator

(OCSI) built in to the device. The OCSI continually monitors

the oxygen output of the device and incorporates a green light

(normal), yellow light (alert), red light (alarm) and a buzzer.

These indicators will all illuminate during the power-on self-test

at start-up. After initially powering ON the device, please allow

up to five (5) minutes for the device to reach its performance

specifications. When only the green indicator light is illuminated,

the device has reached its performance specifications.

Note: If there are three beeps at start-up, this indicates a low battery
condition for the 9-volt battery inside the cabinet of the Eclipse. Please
contact your home care provider and request that the 9-volt battery be
replaced.