Eclipse User Manual — 15

Pulse Dose Mode and autoSat Features

Your Eclipse has a feature, called autoSAT

®

that delivers a

consistent-sized bolus of oxygen up to 40 breaths per minute (see

Max Breath Rate table below). As your breath rate increases, ie.

during ambulation, the autoSAT feature servo-controls the device,

delivering the set bolus size. autoSAT Technology is proprietary

to CAIRE and is utilized during Pulse Mode only.

Mas Breath Rate

Bolus

volume (+/- 15%)

Pulse

Dose

Setting

Bolus Size

(± 15%)

mL

AC Power Supply

& Power Cartridge

(Battery)

DC Power Supply

Max Breath Rate

Max Breath Rate

1.0

16

40

40

2.0

32

40

40

3.0

48

40

40

4.0

64

40

31

5.0

80

37

25

6.0

96

31

20

-

128

23

15

-

160

18

12

-

192

15

10

Step 5: Adjust the Flow Setting to the Prescribed
Level

Using the Flow Setting Buttons, adjust the flow setting to the

prescribed setting supplied by your physician.

Note: Bolus volume decreases as breath rate exceeds published range.
Note: The pulse dose setting (bolus) is not equal to LPM.

WARNING: PULSE DOSE MODE SETTINGS MUST BE DETER-
MINED FOR EACH PATIENT INDIvIDUALLY FOR THEIR NEEDS
AT REST, DURING EXERCISE AND WHEN TRAvELING.

WARNING: THE ECLIPSE MAY NOT BE ABLE TO DETECT ALL INSPIRA-
TORY EFFORTS IN PULSE MODE. IF THE ECLIPSE DOES NOT RESPOND
TO YOUR INSPIRATION EFFORTS, ASK YOUR HOME CARE PROvIDER
TO CHECK THE SENSITIvITY SETTING.

Note: To Equipment Provider: The following oxygen administration acces-
sories are recommended for use with the Oxygen Concentrator:
• Nasal Cannula with 7 feet (2.1 m) of tubing: PN 5408-SEQ
• For any additional recommended accessories, please see the Accessories
Catalog (PN ML-LOX0010) available on www.CAIREmedical.com.

WARNING: IT IS vERY IMPORTANT TO SELECT ONLY THE
PRESCRIBED LEvEL OF OXYGEN. DO NOT CHANGE THE
FLOW SELECTION UNLESS YOU HAvE BEEN DIRECTED TO

DO SO BY A LICENSED CLINICIAN.

Step 6: Begin Using Your Eclipse

Breathe normally through the nasal cannula.

Note: You will achieve longer operating time on the Battery if you operate
your Eclipse in the Pulse Mode (Physician approved mode).

Step 7: Universal Cart Operation

The Universal Cart is designed for use on pedestrian-type

surfaces.

Align the locater holes on the bottom of the device to the tabs on

the Universal Cart.

Align the large threaded screw to the

insert at the back of the Eclipse. Then

hand-tighten the knob on the cart until the

Universal Cart is tightly secured to the

device.

Depress the push-button on the Universal

Cart handle to adjust the height of the

handle.

Step 8: Power Off the Device

Press and hold the “ON/OFF” Button for two (2) seconds to

power OFF your Eclipse.

Store the Eclipse in a cool and dry location for a period no longer

than 1 month. The Eclipse must run for a minimum of 2 hours per

month to ensure proper function of the unit.

Note: DO NOT operate the Eclipse and ambulate while a humidifier is
attached. Remove the humidifier bottle before walking. DO NOT lay the
Eclipse down while attached to a humidifier bottle.