Typical Battery Recharge Time

The typical time to recharge your Battery to achieve 80%

capacity from a fully discharged Battery is between 2 and 5

hours, depending upon the device flow setting. When the battery

is fully charged, the battery gauge will cease cascading. The

battery can be charged in the unit when the Eclipse is turned off

and plugged in.

If the Battery becomes too warm during discharging, recharging

will not begin until the Battery sufficiently cools. Removing the

Battery and allowing it to cool may expedite this cooling process.

The Battery will not recharge when operating from the DC Power

Supply.

The Battery charges when in the device is operating on AC power

only. The Battery will also charge from the optional Desktop

Charger.

Typical New Power Cartridge Operating Times

Flow Rate

Setting

Continuous
Flow (LPM)

mL

SETTING

Pulse Dose Mode

(12 BPM)

0.5

4.4 hours

-

-

1.0

3.7 hours

16mL 1.0

5.4 hours

2.0

2.0 hours

32mL 2.0

5.1 hours

3.0

1.3 hours

48mL 3.0

4.9 hours

64mL 4.0

4.0 hours

80mL 5.0

3.7 hours

96mL 6.0

3.5 hours

128mL

2.5 hours

160mL

2.0 hours

192mL

1.7 hours

Note: Operating the Oxygen Concentrator outside of its normal operating
temperature range can affect performance and decrease battery run time
and/or increase battery charge time. (Refer to the Specifications section in
this manual.) You will achieve longer operating time on the Battery if you
operate your Eclipse in the Pulse Mode (Physician approved mode).