3. removing the temperature and rh sensor cover – Campbell Scientific TurfWeather Weather Station User Manual

Page 50

Advertising
background image

TurfWeather Weather Station

d. Remove the two Phillips head screws on the base of the main body

that hold the mounting cover over the tube that covers the

temperature/relative humidity sensor (FIGURE 6-3).

e. Gently separate these components and pull out the temp/RH sensor

until the connector is exposed.

FIGURE 6-3. Removing the temperature and RH sensor cover

2. Sensor Replacement:

a. Do not contaminate either of the sensor elements by exposing them to

solvents.

b. Unplug the old sensor, and plug in the new sensor.

3. Reassemble the mounting cover and tube assemble with the new sensor

onto the main body.

4. System Test:

a. Connect the weather station to the host computer using the short

RS-232 test cable (if a wired system) and turn on the weather station

power switch. Initiate communications with the weather station using

PC200W.

b. Verify that the temperature and relative humidity sensors are reporting

values. Gently blow air into the sensor protective screen on the

bottom of the weather station and verify that both the temperature and

relative humidity values increase. Stop blowing air on the sensors,

and after a brief period the values should slowly decrease until the

sensors return to equilibrium with the surrounding environment.

c. Reinstall the weather station on its mast, and then reconnect power

and earth ground cables.