18) H/W Config - T1E1 COLD ID

19) VNA PLL Lock Failure Test - Status (01h: Pass, 00h: Fail)

20) VNA PLL Lock Failure Test - Failed data point # (Ignore this byte if the Lock Fail Test Status was Pass)

21) VNA PLL Lock Failure Test - Failed PLL # (Ignore this byte if the Lock Fail Test Status was Pass)

22) VNA Integration Test - Status (01h: Pass, 00h: Fail)

23) VNA Integration Test - Failed data point # (Ignore this byte if the Integration Test Status was Pass)

24) VNA Integration Test - Reserved

25) SPA LO Test - Status (01h: Pass, 00h: Fail, FFh: SPA board not installed)

26) SPA LO Test - Failed data point #

27) SPA LO Test - Failed LO #

28) Module PLD Version

29) Module Attached

30) Module Lock (01h = Locked, 00h = Not Locked)

31) Module Lock Fail Counter (higher byte)

32) Module Lock Fail Counter (lower byte)

33) End of Data (FFh)

S331D + Option 50: 36 bytes

1)

Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2° C) (higher byte)

2)

Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2° C) (lower byte)

3)

Memory Check (01h: Pass, 00h: Fail)

4)

RTC Voltage Check (01h: Pass, 00h: Fail)

5)

Power Voltage in 1/10ths of a Volt (e.g., 124 = 12.4 Volts) (higher byte)

6)

Power Voltage in 1/10ths of a Volt (e.g., 124 = 12.4 Volts) (lower byte)

7)

VNA Lock Failure Counter (higher byte)

8)

VNA Lock Failure Counter (lower byte)

9)

VNA Integrator Failure Counter (higher byte)

10) VNA Integrator Failure Counter (lower byte)

11) SPA LO Failure Counter (higher byte)

12) SPA LO Failure Counter (lower byte)

13) H/W Config - Mother Board ID

14) H/W Config - SPA Board ID

15) H/W Config - T1E1 Board ID

16) H/W Config - PLD1 ID

17) H/W Config - PLD2 ID

18) H/W Config - T1E1 COLD ID

19) VNA PLL Lock Failure Test - Status (01h: Pass, 00h: Fail)

20) VNA PLL Lock Failure Test - Failed data point # (Ignore this byte if the Lock Fail Test Status was Pass)

21) VNA PLL Lock Failure Test - Failed PLL # (Ignore this byte if the Lock Fail Test Status was Pass)

22) VNA Integration Test - Status (01h: Pass, 00h: Fail)

23) VNA Integration Test - Failed data point # (Ignore this byte if the Integration Test Status was Pass)

24) VNA Integration Test - Reserved

25) Status (01h: Pass, 00h: Fail, FFh: T1/E1 board not installed)

26) Carrier Status (01h: carrier present, 00h: No carrier)

27) Frame Sync Status (01h: in frame sync, 00h: Not in frame sync)

28) QRSS pattern sync status (01h: Pattern sync, 00h: Not in sync)

29) QRSS bit error count (01h: Bit error found, 00h: No bit error)

30) T1 - Daly pattern sync status (01h: Pattern sync, 00h: Not in pattern sync)

31) 0 dB CSU Tx Level Check (00h: Pass (> - 2.5 dB), XXh: Value reported by DS2155)

32) -7.5 dB CSU Tx Level Check (00h: Pass (-5.0 to -12.5 dB), XXh: Value reported by DS2155)

33) -15 dB CSU Tx Level Check (00h: Pass (-12.5 to -20.0 dB), XXh: Value reported by DS2155)

34) -22.5 dB CSU Tx Level Check (00h: Pass (-20.0 to -30.0 dB), XXh: Value reported by DS2155)

35) Vpp measurement of 0 dB signal in 1/10ths of a Volt (e.g., 124 = 12.4 Volts)

36) End of Data (FFh)