189) Cable Loss

190) Cable Loss (lowest byte)

191) Average Cable Loss

155

(highest byte)

192) Average Cable Loss

193) Average Cable Loss

194) Average Cable Loss (lowest byte)

195) Status Byte 1: ( 0b = Off , 1b = On)

(LSB)

bit 0 : Marker 1 On/Off

bit 1 : Marker 2 On/Off

bit 2 : Marker 3 On/Off

bit 3 : Marker 4 On/Off

bit 4 : Marker 5 On/Off

bit 5 : Marker 6 On/Off

bits 6-7 : Not Used

196) Status Byte 2: (0b = Off, 1b = On)

(LSB)

bit 0 : Marker 2 Delta On/Off

bit 1 : Marker 3 Delta On/Off

bit 2 : Marker 4 Delta On/Off

bits 3-7 : Not Used

197) Status Byte 3: ( 0b = Off , 1b = On)

(LSB)

bit 0 : Single Limit On/Off

bit 1: CW On/Off

bit 2: Trace Math On/Off

bits 3-5 : Not Used

bit 6 : Limit Type ( 0b = Single; 1b = Multiple)

bit 7 : Unit of Measurement (1b = Metric, 0b = English)

198) Status Byte 4:

(LSB)

bit 0 - 1 : DTF Windowing Mode

bit: 1 0

| |

0 0 - Rectangular (No Windowing)

0 1 - Nominal Side Lobe

1 0 - Low Side Lobe

1 1 - Minimum Side Lobe

bits 2 – 7 : Not Used

199) Status Byte 5 (Cal Status):

00h : Calibration Off

01h : Standard Calibration On

02h : InstaCal Calibration On

03h : Standard FlexCal On

04h : InstaCal FlexCal On

200) VNA Signal Standard

156

(higher byte)

201) VNA Signal Standard (lower byte)

202-324) Not Used

325-1364) Sweep Data (130 points * 8 bytes/point = 1040 bytes)

325-2396) Sweep Data (259 points * 8 bytes/point = 2072 bytes)

325-4460) Sweep Data (517 points * 8 bytes/point = 4136 bytes)

8 bytes for each data point

1. gamma

157

(highest byte)

52

Site Master PM

155 Average Cable Loss is dB * 1000.

156 Index into Standard List (use control byte #89 to retrieve the ASCII string name). “No Standard” sent as FFFEh

157 Gamma data uses 1/10,000 units.