Select Measurement Mode – Control Byte #3 (03h)

Description: Sets the measurement mode of the Site Master. The response byte will not be sent until the mode change is

complete.

See control byte #29 (1Dh) response byte 3 for the current Site Master measurement mode.

Bytes to Follow: 1 byte

1) Measurement Mode

00h: RL Frequency

01h: SWR Frequency

02h: Cable Loss Frequency

10h: RL Distance

11h: SWR Distance

30h: Spectrum Analyzer Mode

31h: Transmission Mode (Option 21 only)

40h: Power Meter Mode (Option 29 only)

60h: T1 Tester Mode (Option 50 only)

70h: E1 Tester Mode (Option 50 only)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error : Invalid measurement mode

238 (EEh) Time-out Error

Set Site Master VNA Scale – Control Byte #4 (04h)

Description: Sets the top and bottom value of the current measurement mode.

Return Loss & Cable Loss:

Unit is dB/1000.

Maximum value sent is 60000 which represents 60.00 dB,

Minimum value sent is 0 which represent 0.00 dB,

Start value < Stop value

SWR:

Unit is 1/1000 (of ratio)

Maximum value sent is 65535 which represents 65.53

Minimum value sent is 1000 which represents 1.00

Start value < Stop value

See control byte #29 (1Dh) response bytes 36 to 43 for current Site Master scaling.

Bytes to Follow: 8 bytes

1)

Scale Start (highest byte)

2)

Scale Start

3)

Scale Start

4)

Scale Start (lowest byte)

5)

Scale Stop (highest byte)

6)

Scale Stop

7)

Scale Stop

8)

Scale Stop (lowest byte)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error : Invalid scale range

238 (EEh) Time-out Error