Save system setup œ control byte #18 (12h) 20, Save system setup – control byte #18 (12h) – Anritsu Site Master S331D User Manual

Page 26

Advertising
background image

300) Status Byte 5

( 0b = Off/Beep if data is below line, 1b = On/Beep if data is above line)

(LSB)

bit 0 : Multiple Limit Upper Segment 3 Status On/Off

bit 1 : Multiple Limit Upper Segment 3 Beep Level ABOVE/BELOW

bit 2 : Multiple Limit Upper Segment 4 Status On/Off

bit 3 : Multiple Limit Upper Segment 4 Beep Level ABOVE/BELOW

bit 4 : Multiple Limit Upper Segment 5 Status On/Off

bit 5 : Multiple Limit Upper Segment 5 Beep Level ABOVE/BELOW

bit 6 : Multiple Limit Lower Segment 1 Status On/Off

bit 7 : Multiple Limit Lower Segment 1 Beep Level ABOVE/BELOW

28

303) Status Byte 6

(0b = Off/Beep if data is BELOW line, 1b = On/Beep if data is ABOVE line)

(LSB)

bit 0 : Multiple Limit Lower Segment 2 Status On/Off

bit 1 : Multiple Limit Lower Segment 2 Beep Level ABOVE/BELOW

bit 2 : Multiple Limit Lower Segment 3 Status On/Off

bit 3 : Multiple Limit Lower Segment 3 Beep Level ABOVE/BELOW

bit 4 : Multiple Limit Lower Segment 4 Status On/Off

bit 5 : Multiple Limit Lower Segment 4 Beep Level ABOVE/BELOW

bit 6 : Multiple Limit Lower Segment 5 Status On/Off

bit 7 : Multiple Limit Lower Segment 5 Beep Level ABOVE/BELOW

304)

Status Byte 7

bits 0-6: Number of sweeps to average (1-25, 1 implies no averaging)

bit 7: Not Used

305)

Reference Level Offset

29

(highest byte)

306)

Reference Level Offset

307)

Reference Level Offset

308)

Reference Level Offset (lowest byte)

309-338) Not Used

339-1938) Sweep Data

(400 points * 4 bytes/point= 1600 bytes)

4 bytes for each data point

1. dBm

30

MSB

2. dBm

3. dBm

4. dBm LSB

Site Master Returns (For invalid sweeps/empty stored sweep locations): 11 bytes

1-2) Number of following bytes (9 bytes for invalid sweep recall)

3-4) Model # (unsigned integer, 14h for Site Master S33xD)

5-11) Extended Model # (7 bytes in ASCII)

Site Master Returns (Invalid sweep location): 1 byte

1) 224 (E0) Parameter Error: Invalid sweep location

Save System Setup – Control Byte #18 (12h)

Description: Saves current system setup parameters to a specific setup store location.

The Site Master saves all parameters described in Query System Status - control byte #29 (1Dh), (except Serial Port Echo

Status) to the specified store location. Store location 0 is the run-time setup of the Site Master. It holds the power-on defaults

of the Site Master.

20

Site Master PM

28 LOWER limits always trigger an error beep if data is BELOW the limit segment, for example, this bit is always 0b.

29 Value sent as ( value in dBm * 1000 ) + 270,000

30 Value sent as ( value in dBm * 1000 ) + 270,000