Anritsu Site Master S331D User Manual

Page 79

Advertising
background image

Power Meter Relative Mode On/Off – Control Byte #41 (29h)

(Option 29 only)

Description: Enable or disable Power Meter Relative Mode.

Bytes to Follow: 1 byte

1)

Relative Mode State

00h = Off

01h = On w/ trigger (use the current power level as a reference power level)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid parameter

238 (EEh) Time-out Error

Power Meter Offset Mode On/Off – Control Byte #42 (2Ah)

(Option 29 only)

Description: Enable or disable Power Meter Offset Mode.

Bytes to Follow: 5 bytes

1) On/Off (01h = On, 00h = Off)

2 - 5) Offset Power level in dB (Multiplied by 1000)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid parameter

238 (EEh) Time-out Error

Note:If you turn the Offset mode off, you must still send the other bytes. Bytes 2 - 5 will be ignored.

Power Meter Zero Mode On/Off – Control Byte #43 (2Bh)

(Option 29 only)

Description: Enable or disable Power Meter Zeroing Mode.

Bytes to Follow: 1 byte

1)

Zero Mode Status

00h = Off

01h = On with trigger (current power level is referenced as -80 dBm)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid status

238 (EEh) Time-out Error

Power Meter RMS Averaging On/Off – Control Byte #44 (2Ch)

(Option 29 only)

Description: Set Power Meter RMS Averaging. Enabling to one of three different levels, or off.

Bytes to Follow: 1 byte

1)

RMS Averaging State

00h = Off

01h = On (Low) with trigger (current power level is referenced as -80 dBm)

02h = On (Medium)

03h = On (High)

Site Master Returns: 1 byte

Site Master PM

73