3)

Ref Level

4)

Ref Level (lowest byte)

5)

dB/div (highest byte)

6)

dB/div

7)

dB/div

8)

dB/div (lowest byte)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid scale

238 (EEh) Time Out Error

Notes:

Ref Level is sent as the (Ref Level * 1000) + 270,000 (0 dBm = 270,000, 20 dBm = 290000, -120 dBm = 150,000)

Scale should be sent as (dBm * 1000) (e.g., -12.34 dBm = -12340)

Set Spectrum Analyzer Marker – Control Byte #102 (66h)

Description: Sets an individual Spectrum Analyzer marker.

Bytes to Follow: 5 bytes

1)

Marker Number (01h = marker 1, 02h = marker 2, 03h = marker 3, 04h = marker 4, 05h = marker 5, 06h

= marker 6)

2)

Marker Line On/Off (01h = On, 00h = Off)

3)

Marker Delta Status On/Off (01h = On, 00h = Off)

4)

Marker Value (higher byte)

5)

Marker Value (lower byte)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid marker number, status or position

238 (EEh) Time Out Error

Note: Marker Value is between 0 and 400, inclusive: Point = (400 * (marker freq - start freq)) / span

Set Spectrum Analyzer Single Limit – Control Byte #103 (67h)

Description: Sets the position and On/Off Status of the Limit Line.

Bytes to Follow: 6 bytes

1)

Limit Number (1 for Site Master)

2)

Limit Line On/Off (01h = On, 00h = Off)

3)

Beep at Limit On/Off (01h = On, 00h = Off)

4)

Limit Value (highest byte)

5)

Limit Value

6)

Limit Value

7)

Limit Value (lowest byte)

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid limit number, status or value

238 (EEh) Time Out Error

Note: Limit Value is sent as the (Limit Value * 1000) + 270,000 (0 dBm=270,000, 20 dBm=290000, -120 dBm=150,000)