Set Spectrum Analyzer Peak Hold – Control Byte #105 (69h)

Description: Sets the max hold and min hold settings on the Spectrum Analyzer.

Bytes to Follow: 1 byte

1)

Peak Hold State

00h – Peak Hold Off

01h – Max Hold On

02h – Min Hold On

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid state

238 (EEh) Time Out Error

Set Spectrum Analyzer Resolution Bandwidth – Control Byte #106 (6Ah)

NOTE: This command exists for backward compatibility with the S33xC models. Features new to the

S33xD models are not available here. To access the new features use Control Byte #141 (8Dh).

Description: Sets the resolution BW frequency for the Spectrum Analyzer.

Bytes to Follow: 1 byte

1)

Resolution Bandwidth Index

00h – 10 kHz BW

01h – 30 kHz BW

02h – 100 kHz BW

03h – 1 MHz BW

Site Master Returns: 1 byte

1)

255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid RBW Index

238 (EEh) Time Out Error

Set Spectrum Analyzer Video Bandwidth – Control Byte #107 (6Bh)

NOTE: This command exists for backward compatibility with the S33xC models. Features new to the

S33xD models are not available here. To access the new features use Control Byte #142 (8Eh).

Description: Sets the video BW frequency for the Spectrum Analyzer.

Bytes to Follow: 1 byte

1)

Video Bandwidth Index

00h – 100 Hz BW

01h – 300 Hz BW

02h – 1 kHz BW

03h – 3 kHz BW

04h – 10 kHz BW

05h – 30 kHz BW

06h – 100 kHz BW

07h – 300 kHz BW

Site Master Returns: 1 byte