Standard Commands for Programmable Instruments (SCPI)

103

Trigger Subsystem

This subsystem controls remote triggering of the power supply.

TRIG[:IMM]

When the trigger subsystem is armed,

TRIG

generates a trigger signal. The trigger will then:

1. Initiate the output change from the pending level to the immediate level (see

CURR[:LEV]:TRIG

and

VOLT[:LEV]:TRIG

).

2. Clear the WTG bit in the Status Operation Condition register.
3. If

INIT:CONT

has been sent, the trigger subsystem is immediately rearmed for subsequent triggers. This means that as

soon as it is cleared, the WTG bit is again set to 1.

Command Syntax

TRIG

er

[:IMM

ediate

]

Examples

TRIG TRIGGER:IMMEDIATE

TRIG:SOUR

Sets the trigger source to bus commands. Since the power supply has no other trigger source than the GP-IB bus, this
command need not be used. It is included in the command set to provide programming compatibility with other instruments
(such as the Agilent Electronic Load family) that may have more than one trigger source.

Command Syntax

TRIG

er:

SOUR

ce

BUS

Parameters

<CRD>

BUS

*RST

Value

BUS

Examples

TRIG: SOUR BUS TRIGGER: SOURCE BUS

Query Syntax

TRIG:SOUR?

Returned Parameters

<CRD>

BUS

Voltage Subsystem

This subsystem programs the output voltage of the power supply.

VOLT[:LEV]

Sets the immediate voltage level or the pending triggered voltage level of the power supply. The immediate level is the
voltage programmed for the output terminals. The pending triggered voltage level is a stored value that will be transferred to
the output when a trigger occurs. If no pending triggered level has been programmed, then the immediate command
programs both levels to the same value. Once a triggered level has been programmed, it is unaffected by subsequent changes
to the immediate level.

Command Syntax

[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude] <NRf+>
[SOURce]:VOLTage[:lEVel]:TRIGgered[:AMPlitude] <Nrf+>

Parameters

See "VSET" in Table 3-7 of Section III.

Default Sufflx

V

Examples

VOLT 200 MV VOLTAGE:LEVEL 200 MV
VOLTAGE:LEVEL:IMMEDIATE:AMPLITUDE 2.5
VOLT:TRIG 20 VOLTAGE:LEVEL:TRIGGERED 20
VOLTAGE:LEVEL:TRIGGERED:AMPLITUDE 5.5