Programming the Agilent 603xA Power Supplies Using Basic

117

120

!

130

DEF FNSettled(OPTIONAL Band, Rdgs.Timelimit)

140 !

BAND=SETTLING BAND IN VOLTS

150 !

RDGS=READINGS WITHIN BAND

160 !

TIMELIMIT=SETTLING TIME ALLOWED, SECONDS

170

COM /Ps/ @Ps

180

OUTPUT @Ps:”VSET?”

190

ENTER @Ps;Vset

200

SET TIMEDATE 2.1E+11

210

!

220

Band_p=.01

230

Rdgs_p=2

240

Timelimit_p=5

250

Counter=0

260

!

270

IF NPAR>0 THEN Band_p=Band

280

IF NPAR>l THEN Rdgs_p=Rdgs

290

IF NPAR>2 THEN Timelimit_p=Timelimit

300

!

310

WHILE FNElapsed_time<Timelimit_p

320

IF ABS(FNVout-Vset)<Band_p THEN

330

Counter=Counter+l

340

IF Counter>=Rdgs p THEN RETURN l

350

ELSE

360

Counter=0

370

END IF

380

END WHILE

390

!

400

RETURN 0

410

FNEND

420

!

430

!

440

DEF FNVout

450

COM /Ps/ @Ps

460

OUTPUT @Ps; “VOUT?”

470

ENTER @Ps;Vout

480

RETURN Vout

490

FNEND

500

!

510

!

520

DEF FNElapsed_time

530

RETURN TIMEDATE-2.1E+11

540

FNEND

Explanation:

10:

Assign I/O path name for power supply

20:

Declare common block for I/O path name

30:

Clear power supply, set voltage to 20 Volts, current limit to 2 Amps

50:

Downprogram power supply to 100 mV

60:

Format statement for display

70,90:

If P.S. output has settled to within 20 mV of programmed value (for 5 consecutive readings) print message

130:

Define FN settled. (optional parameters described in lines 140-160)

170:

Bring in common block for I/O path name

180,190:

Read set value for voltage into VSET

200:

Initialize real time clock in computer

220-250:

Assign default values to variables corresponding to elements in parameter list. (Cannot

reference optional parameters in context if not passed in parameter list.)

270-290:

If parameters are passed, assign values to corresponding subprogram variables

310-380:

Repeat 320-370 until time limit is reached

320-330:

If absolute value of output voltage (-Set voltage is within settling band) increment

counter

340:

If counter=Desired number of readings within band, return 1

360:

If voltage left setting band, reset counter

400:

If time has elapsed, return 0

440-490:

Function reads and returns output voltage

520-540:

Function returns time elapsed since real time clock was initialized on line 200

Advertising