Accumulated status register mask & fault registers – Agilent Technologies 6028A User Manual

Page 61

Advertising
background image

Operating Instructions

61

For example, if bits for both ERR (128) and CC (2) are set, the power supply would send ASCII digits 1 3 0 (128 +2 =
130).

Bits remain set in the status register as long as the corresponding conditions are true.

Table 3-9. Status Register

Bit Position

8

7

6

5

4

3

2

1

0

Bit Weight

256

128

64

32

16

8

4

2

1

Condition

RI

ERR

FOLD

AC

OT

OV

OR

CC

CV

CV Constant Voltage Mode
CC Constant Current Mode
OR Overrange
OV Overvoltage Protection Circuit Tripped
OT Overtemperature Protection Circuit Tripped
AC AC Line Dropout or Out of Range
FOLD Foldback Protection Circuit Tripped
ERR Remote Programming Error
RI Remote Inhibit (INH)

Accumulated Status Register

. Reading the status register provides the controller only the state of the power supply at the

time STS? was received. A condition that lasts only momentarily may not be observed even with frequent polling of the
status register. To ensure that a temporary change can be noted by the controller, the power supply maintains an
accumulated status (astatus) register. Table 3-9 describes the astatus register as well as the status register. A bit in the astatus
register will be 1 if the corresponding bit in the status register has been 1 at any time since the astatus register was last read.

The astatus register can be read by sending:

ASTS?

and addressing the power supply to talk. The response from the power supply is in this format:

ASTS xxx

where xxx is decoded the same way as in the status register readback.

The astatus register is reset to the present value of the status register immediately after it is read by the ASTS?
query.

Mask and Fault Registers

. The power supply has two additional registers, the mask register and the fault register, both of

which are arranged like the status register (Table 3-9). The mask register is maintained by the user, and is used to specify
which bits in the status register are enabled (unmasked) to set bits in the fault register. A bit is set in the fault register when
the corresponding bit in the status register changes from 0 to 1 and the corresponding bit in the mask register is 1. Whenever
any bit is set in the fault register the FAU bit is set in the serial poll register.