Table C-5. Summary of Error Messages (continued)

Error Number

Error String [Description/Explanation/Examples]

-178

Expression data not allowed

-180

Macro error

-181

Invalid outside macro definition [e.g., ’$1’ outside macro definition.]

-183

Invalid inside macro definition

-184

Macro parameter error

- 200

Execution error [generic]

- 220

Parameter error

- 221

Settings conflict [uncoupled parameters]

- 222

Data out of range [e.g., too large for this instrument]

- 223

Too much data [out of memory; block, string, or expression too long]

- 240

Hardware error

- 241

Hardware missing

- 270

Macro error

- 272

Macro execution error

- 273

Illegal macro label

- 276

Macro recursion error

- 277

Macro redefinition not allowed

- 310

System error

- 313

Calibration memory lost [out of cal due to memory failure]

-330

Self-test failed [more specific data after ";’’]

- 350

Too many errors [error queue overflow]

- 400

Query error

- 410

Query INTERRUPTED [query followed by DAB or GET before response complete]

- 420

Query UNTERMINATED [addressed to talk, incomplete programming message
received]

- 430

Query DEADLOCKED [too many queries in command string]

- 440

Query UNTERMINATED [after indefinite response]

SCPI COMMAND SUMMARY

This summary lists all power supply subsystem commands in alphabetical order, followed by all common commands in
alphabetical order. See Table 3-7 in Section III for the numerical parameters for each model of the Agilent Series 603xA
power supplies. Table C-6 is a cross reference to the corresponding ARPS commands in Table 3-7.

Command Summary

Subsystem Commands

Command

Parameters

ABOR

(none)

[SOUR]:CURR[:LEV][:IMM][:AMPL]

<space> <NRf+ >[suffix]

[SOUR]:CURR?

(none) or <space>MlN or <space>MAX

[SOUR]:CURR[:LEV]:TRIG[:AMPL]

<space> <NRf+ >[suffix]

[SOUR]:CURR[:LEV]:TRIG?

(none) or <space>MlN or <space>MAX

[SOUR]:CURR:PROT:STAT

<space>0 or OFF <space>1 or ON

[SOURl:CURR:PROT:STAT?

(none)

INIT[:IMM]

(none)

INIT:CONT

< space > 0 or OFF or < space > ON

INIT:CONT?

(none)

MEAS:CURR[:DC]?

(none)

MEAS:VOLT[:DC]?

(none)

Advertising