Table 1-2. Supplemental Characteristics (continued)

DC Floating Voltage

: Either output terminal may be floated up to the following voltage (including the output voltage) from

earth ground:

±

240 Vdc on Models 6031A, 6032A, 6033A, and 6038A

±

550 Vdc on Models 6030A and 6035A

Exceeding these voltage can result in damage to the equipment.

Remote Sensing

: The power supply maintains specifications at the load with up to 0.5 volt drop per load lead. Operation

with up to 2 volts per load lead is possible with some degradation of the load effect specification.

Absolute maximum ratings

between FLT Terminals 1 and 2 is 17.5 Volts dc. (external resistor required to limit current to 1.25 mA max)

between INH Terminals 3 and 4 is 5.5 Volts dc.

between Terminals 1 or 2 to 4 or chassis to ground is 42 Volts dc.

Fault (FLT) levels (Terminals 1 and 2) (All models)

Ioh (Low Level Output Current) is + 1.2 mA maximum

Vol (Low Level Output Voltage) is 0.5 Volts Maximum

Inhibit (INH) Levels (Terminals 3 and 4) (All models)

Vih (High Level Input Voltage) is 2 Volts

Vil (Low Level Input Voltage) 0.5 Volts Maximum

tw (Pulse Width) (minimum) = 50 ms (2 ms typ)

td (Delay Time) = 2 ms typ

Iil (Low Level Input Current) = -1.25 mA maximum

Multiple Operations

: Up to two similar units may be connected in series or auto-parallel, to provide increased output

capabilities. Mixing supplies with dissimilar output capabilities is not recommended because under certain conditions, the
lower output supply may be stressed beyond its maximum voltage and or current capabilities by the higher output supply.

Temperature Rating (

°

C):

Operating is 0-50 (Agilent 6030/6031/6032/6035); 0-55 (Agilent 6033/6038)

Storage is - 40 + 75 (all models)

Weight kg. (Ibs)

Model

Agilent 6030A

Agilent 6031A

Agilent6032A

Agilent 6033A Agilent 6035A

Agilent 6038A

Net

16.3 (36)

17.2 (38)

16.3 (36)

9.6 (21)

16.3 (36)

9.6 (21)

Shipping

21.7 (48)

22.7 (50)

21.8 (48)

11.4 (25)

21.7 (48)

11.4 (25)

Dimensions

: See Figure 2-1.

Certification

:

The unit is designed to comply with these requirements:

ICE 348-Safety Requirements for Electronic Measuring Apparatus.

CSA Electrical Bulletin 556B-Electronic Instruments and Scientific Apparatus for Special Use and Applications.

VDE 0871.6.78 Level B-RFI Suppression of Radio Frequency Equipment for Industrial, Scientific, and Medical (ISM)
and similar purposes.

VDE 0411-Electronic Measuring Instruments and Automatic Controls.

UL 1244-Electrical and Electronic Measuring & Testing Equipment.

ANSI C39.5 Part 0 Draft 8-Electrical Testing, Measurement, and Control Equipment.