Standard Commands for Programmable Instruments (SCPI)

96

Current Subsystem

This subsystem programs the output current of the power supply.

CURR[:LEV]

Sets the immediate current level or the pending triggered current level of the power supply. The immediate level is the
current programmed for the output terminals. The pending triggered current level is a stored value that will be transferred to
the output when a trigger occurs. If no pending triggered level has been programmed, then the immediate command
programs both levels to the same value. Once a trigger level is programmed, it is unaffected by subsequent changes to the
immediate level.

Command Syntax

[

SOUR

ce]:

CURR

ent[:

LEV

el] [:

IMM

ediate][:

AMP

Litude]

<NRf+>

[

SOUR

ce][:

CURR

ent[:

LEV

el]:

TRIG

gered [:

AMPL

itude] <

NRf+>

Parameters

See "ISET" in Table 3-7 of Section III.

Default Sufflx

A

Examples

CURR 200 MA CURRENT:LEVEL 200 MA
CURRENT:LEVEL:IMMEDIATE:AMPLITUDE2.5
CURR:TRIG 20 CURRENT:LEVEL:TRIGGERED 20

Query Syntax

[

SOUR

ce]:

CURR

ent[:

LEV

el] [:

IMM

ediate][:

AMP

Litude]?

[

SOUR

ce]:

CURR

ent[:

LEV

el] [:

IMM

ediate][:

AMPL

itude]?

MAX

[

SOUR

ce]:

CURR

ent[:

LEV

el] [:

IMM

ediate][:

AMPL

itude]?

MIN

[

SOUR

ce]:

CURR

ent[

LEV

el]:

TRIG

ger [:

AMPL

itude]?

[

SOUR

ce]:

CURR

ent[

LEV

el]:

TRIG

ger [:

AMPL

itude]?

MAX

[

SOUR

cel:

CURR

ent[:

LEV

ell:

TRIG

ger [:

AMPL

itude]?

MIN

Returned Parameters <NR3> CURR?

and

CURR:TRIG?

return presently programmed current levels. If the

TRIG

level is

not programmed, both returned values are the same.

CURR? MAX

and

CURR? MIN

return the maximum and minimum programmable immediate current levels.

CURR:TRIG? MAX

and

CURR:TRIG?

MIN return the maximum and minimum programmable triggered current levels.

CURR:PROT:STAT

Enables or disables the power supply overcurrent protection function (OCP). If the OCP function is enabled and the output
current exceeds the programmed immediate level then:

the power supply output is disabled

the Questionable Condition status register OC bit is set (see "Status Reporting").

An overcurrent condition can be cleared with the

OUTP:PROT:CLE

command after the condition that caused the OCP trip

is removed.

Command Syntax

[

SOUR

ce][:

CURR

ent:

PROT

ection:

STATe

<bool>

Parameters

0

or

OFF

1

or

ON

*RST

Value

OFF

Examples

CURR: PROT: STAT 0

CURRENT: PROTECTION: STATE OFF

CURR:PROT:STAT 1

CURRENT:PROTECTION:STATE ON

Query Syntax

[SOURce]:CURRent:PROTection:STATe?

Returned Parameters

<NRl> 0 or 1