1 dummy terminal voltage switching, When only clock 1 output is used – Anritsu MP1763C User Manual

Page 48

Advertising
background image

4-14

Section 4 OPERATING INSTRUCTIONS

4.3.1 DUMMY terminal voltage switching

When only one of the CLOCK 1 and CLOCK 1 complemen-

tary outputs is used, use the attached semirigid cable (10 cm)

to connect the unused side to the DUMMY terminal.

If the termination condition of the side used is 50

, and

GND, press the ECL key while holding down the GUARD

key of

to turn on the 50

GND lamp and set the DUMMY

terminal voltage to GND.

If the termination condition of the side used is ECL (50

,

-2 V), press the ECL key while holding down the GUARD

key of

to turn on the ECL lamp and set the DUMMY termi-

nal voltage to -2 V.

Notes:

1. When both outputs, match their termination

conditions. (Setting one to 50

, GND termina-

tion and the other to ECL termination is not al-

lowed.)

2. When only one output is used with 50

, and

GND termination, the unused output can also

be connected to a 50

terminator instead of

DUMMY. However, only when one output is

used with ECL termination, connect the unused

output to DUMMY.

3. When the termination condition is switched

from 50

Ω GND to ECL, amplitude 0.8Vp-p

and offset -0.9V (V

OH

) are automatically set.

4. Since the DATA and DATA outputs are in

dependent, connection to the DUMMY termi-

nal is unnecessary.

5. If a commercially available ECL terminator is

used to measure the output waveform, wave

form distortion (ringing) may be observed.

This phenomenon depends on the characteris-

tics of the ECL terminator and does not mean

that the output of this equipment contains

waveform distortion.

When only CLOCK 1 output is used

OUTPUT

ps

DELAY TIME

BUSY

CLOCK 1

50

GND

ECL

AMPLITUDE

Vp - p

OFFSET

V

ECL

GUARD

ECL

DISPLAY

DATA/DATA
TRACKING

GUARD

OUTPUT

ON

OFF

OFFSET

V

OH

V

TH

V

OL

DISPLAY

1/1 SPEED

1/4 SPEED
(OPTION.03)

DATA

50

GND

ECL

DATA

DATA

AMPLITUDE

Vp - p

OFFSET

V

CLOCK 1

When only CLOCK 1 output is used

Advertising