2 amplitude, offset, and delay setting, Output ‘high level’ is made the standard. v, If the offset is changed to 1 v – Anritsu MP1763C User Manual

Page 49

Advertising
background image

4-15

4.3.2 Amplitude, offset, and delay setting

When you want to set front panel DATA/DATA output

and CLOCK 1/CLOCK 1 output, switch to 1/1 SPEED.

When you want to set rear panel 1/4 SPEED OUTPUT

(OPTION), switch to 1/4 SPEED.

Switches whether the DATA output or DATA output is

set. When DATA/DATA TRACKING is turned on, the

DATA output amplitude and offset become the same as

the DATA output set value.

Set Amplitude to the desired value by turning the rotary

encoder.

Set the offset standard value.

V

OH

:

Output ‘High Level’ is made the standard.

V

TH

:

The output level center value is made the standard.

V

OL

:

Output “Low Level” is made the standard.

Set the offset value. This value shows the value of the ref-

erence specified in

.

The waveform is as follows when the amplitude is 1 Vp-p

and the offset is 1 V

OL

.

If the offset is changed to 1 V

TH

in the above condition:

Set CLOCK in the same way. The offset reference is

shared by DATA and CLOCK.

4.3 Output Interface

OUTPUT

ps

DELAY TIME

BUSY

ECL

DISPLAY

DATA/DATA
TRACKING

GUARD

OUTPUT

ON

OFF

OFFSET

V

OH

V

TH

V

OL

DISPLAY

1/1 SPEED

1/4 SPEED
(OPTION.03)

DATA

50

GND

ECL

DATA

DATA

AMPLITUDE

Vp - p

OFFSET

V

CLOCK 1

V

V

V

OH

TH

OL

1Vp-p

1V

OL

1.5V

TH

2V

OH

1Vp-p

1V

TH

1.5V

OH

0.5V

OL

Advertising