• Place a .001”-.003” shim between the leveling set screws
and mounting surface.
• Adjust each set screw until a slight drag is felt on the shim.
• Evenly tighten the 8-32 (M4) reading head mounting
screws.
11
Acu-Rite Companies Inc.
ENC 125 T/E
Encoder Installation Procedure
Leveling screw (3)
• Slide the brackets away from the reading head.
• Remove the alignment brackets and save for future use.
Alignment bracket
removal
Twist 45°
Slide brackets away from reading head and cable.
• Route the cables with slack loops to allow for axis motion.
• Secure excess cable by fastening with clips or ties.
• Attach the linear encoder connectors to the readout.
Connecting
Secure excess
cable
Provide slack
loops
Attach connector