Checking the Installation

ENC 125 T/E

• Zero the display and indicator.
• Move axis to the end of it’s travel return to dial zero.
• Readout should read zero ± 1 count.

Place dial indicator at the end of the moving

component (scale assembly or reading head).

Repeatability
Test

• Move the axis and compare the display to the movement.

Readout

Counting Test

Counting Test:
• Configure the readout’s encoder and display resolution (see

manual).

• Move the axis and compare the display to the movement.

Repeatability Test:
• Locate an indicator on one end of the encoder and zero the

readout and indicator.

• Move the axis through the full travel and return to dial zero.
• Readout should read zero ± 1 count.

These steps will confirm proper operation of your installation.
The counting Test confirms proper electrical operation. The
Repeatability Test checks the installation integrity