ACU-RITE ENC 125 T/E User Manual

Page 16

Advertising
background image

Acu-Rite Companies Inc.

14

Trouble Shooting

ENC 125 T/E

If you experience difficulties with your installation, do the
following to determine the problem.

Checking the Readout

Difficulties on more than one axis are usually associated with
the readout. Follow these steps to determine if your
difficulties are associated with the readout:

• Insure that the linear encoder connectors are correctly

seated.

• Swap linear encoder cables at the readout to see if the

problem is still shown in the same display.

• If the problem remains in the same display, the readout is

in error.

• If the problem follows the connection change, the linear

encoder may be in error.

If the Readout is at fault, refer to “What to do” to arrange for
the parts necessary to repair your system. If a linear encoder
appears to be at fault, proceed with “Checking the Linear
Encoders
”.

Checking the Linear Encoders

Problems on a single axis are usually associated with the
linear encoder or its installation. Difficulties can be caused
by improper installation, loose or misaligned bracketry, or a
damaged or inoperable encoder.

Follow these steps to determine the cause of your system
difficulties:
• Confirm that your bracketry and installation does not

interfere with other machine structures through the entire
length of the linear encoder travel.

• Check for loose fasteners. If you find loose fasteners, first

confirm that the linear encoder is installed to the
tolerances specified and then retighten the fasteners as
required.

• Confirm that the linear encoder is installed to the required

tolerances by checking the alignment tolerances specified
on Page 4. If the installation does not meet the tolerances,
reinstall the encoder according to the “Installation
Procedure
”.

• Perform a Repeatability Test as described on Page 11. If

the linear encoder is installed to the required tolerances,
the bracketry and encoder have been checked for
interferences and loose fasteners, and the encoder fails the
repeatability test, the encoder is likely at fault.

Do not attempt to repair the reading head or scale assembly.
The ENC 125 is field serviceable by assembly replacement
only. Attempts to repair the encoder can permanently
damage it and void the warranty.

What to do
If an Acu-Rite Companies Inc. linear encoder or readout is
found to be at fault, please contact your authorized
Distributor for instructions prior to removing the encoders or
readout .

Advertising